Development of two-channel prototype ITER vacuum ultraviolet spectrometer with back-illuminated CCD and micro-channel plate detectors
Development of two-channel prototype ITER vacuum ultraviolet spectrometer with back-illuminated CCD and micro-channel plate detectors
A vacuum ultraviolet (VUV) spectrometer of a five-channel spectral system is designed for ITER main plasma impurity measurement. To develop and verify the system design, a two-channel prototype system is fabricated with No. 3 (14.4-31.8 nm) and No. 4 (29.0-60.0 nm) among the five channels. The optic...
Saved in:
Personal Name(s): | Seon, C.R. |
---|---|
Choi, S.H. / Cheon, M.S. / Pak, S. / Lee, H.G. / Biel, W. / Barnsley, R. | |
Contributing Institute: |
Plasmaphysik; IEK-4 |
Published in: | Review of scientific instruments, 81 (2010) S. 10E508 |
Imprint: |
[S.l.]
American Institute of Physics
2010
|
Physical Description: |
10E508 |
DOI: |
10.1063/1.3478550 |
PubMed ID: |
21034036 |
Document Type: |
Journal Article |
Research Program: |
Fusion |
Series Title: |
Review of Scientific Instruments
81 |
Subject (ZB): | |
Link: |
Get full text OpenAccess OpenAccess |
Publikationsportal JuSER |
Please use the identifier: http://hdl.handle.net/2128/16766 in citations.
A vacuum ultraviolet (VUV) spectrometer of a five-channel spectral system is designed for ITER main plasma impurity measurement. To develop and verify the system design, a two-channel prototype system is fabricated with No. 3 (14.4-31.8 nm) and No. 4 (29.0-60.0 nm) among the five channels. The optical system consists of a collimating mirror to collect the light from source to slit, two holographic diffraction gratings with toroidal geometry, and two different electronic detectors. For the test of the prototype system, a hollow cathode lamp is used as a light source. To find the appropriate detector for ITER VUV system, two kinds of detectors of the back-illuminated charge-coupled device and the microchannel plate electron multiplier are tested, and their performance has been investigated. |