This title appears in the Scientific Report :
2014
Please use the identifier:
http://dx.doi.org/10.1016/j.nima.2014.01.007 in citations.
Absolute efficiency measurements with the 10B based Jalousie detector
Absolute efficiency measurements with the 10B based Jalousie detector
The 10B based Jalousie detector is a replacement for 3He counter tubes, which are not affordable any more for large area detectors due to the 3He crisis. In this paper we will show absolute detection efficiency measurements, conducted at the HEiDi instrument at the FRM II in Munich, which prove that...
Saved in:
Personal Name(s): | Modzel, G. (Corresponding Author) |
---|---|
Henske, M. / Houben, Andreas / Köhli, M. / Lennert, P. / Meven, M. / Schmidt, C. J. / Schmidt, U. / Schweika, W. | |
Contributing Institute: |
JARA-FIT; JARA-FIT Streumethoden; PGI-4 Streumethoden; JCNS-2 JCNS-FRM-II; JCNS-FRM-II |
Published in: | Nuclear instruments & methods in physics research / A, 743 (2014) S. 90 - 95 |
Imprint: |
Amsterdam
North-Holland Publ. Co.
2014
|
DOI: |
10.1016/j.nima.2014.01.007 |
Document Type: |
Journal Article |
Research Program: |
JCNS |
Subject (ZB): | |
Publikationsportal JuSER |
The 10B based Jalousie detector is a replacement for 3He counter tubes, which are not affordable any more for large area detectors due to the 3He crisis. In this paper we will show absolute detection efficiency measurements, conducted at the HEiDi instrument at the FRM II in Munich, which prove that the predicted efficiency can be achieved. This validates the detector concept and its adoption in the POWTEX instrument. |